CMU MEMS Laboratory Publication Abstract

 

in Proceedings of the 2nd Annual International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine & Biology (EMBS-MMB), May 2-4, 2002, Madison, Wisconsin.
Endoscopic optical coherence tomography with a micromachined mirror
T. Xie, H. Xie, G. Fedder and Y. Pan
ABSTRACT:
This paper presents experimental results of an endoscopic optical tomographic (EOCT) imaging system based on a micromachined mirror. Due to the use of the micromachined mirror, the endoscopic OCT system is highly suitable for non-invasive imaging diagnosis of a wide variety of inner organs. Ex-vivo imaging of rabbit bladder was performed. The microstructures of the bladder walls could be delineated with a traverse resolution of 20 µm and an axial resolution of 10 µm. The results have demonstrated the potential of EOCT for early diagnosis of bladder cancer and precise guiding of excisional biopsy.
© 2002 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.
Full paper (PDF) (opens in new window).


This page was generated in 0.014126 seconds at 02:29:12 pm UTC on 28 Mar 2024.